Automated Dynamic
Headspace (DHS) is available for
the GERSTEL MultiPurpose Sampler (MPS). The DHS station
efficiently extracts and concentrates VOCs from liquid or solid
samples placed in standard Headspace vials. The sample is
thermostated and agitated while the headspace is purged with a
controlled flow of inert gas, providing controlled and efficient
extraction conditions. Sample temperatures can be selected from
10 °C to 200 °C. Lowering the temperature provides a means of
reducing the amount of water vapor released and ultimately
retrapped. The DHS station additionally holds a replaceable
adsorbent- or sorbent filled tube used for analyte concentration.
The temperature of the adsorbent tube during the DHS process can
be varied from 20 °C to 70 °C for optimal trapping of the
analytes in question.
Adsorbent tubes used
are standard GERSTEL Thermal
Desorption Unit (TDU) tubes. Following analyte concentration,
the tube is automatically transferred to the TDU for Thermal
Desorption and analyte transfer to the GC. Tubes are fitted with
individual adapters enabling both automated movement and
leak-free sealing during the DHS and Thermal Desorption steps
and during storage in the autosampler tray.
Automated DHS combined
with sample introduction to a GC
or GC/MS system is performed by the MPS and TDU. Up to 98
samples can be processed automatically in one sequence.
Parallel processing
of samples enable the DHS process
to be performed during the chromatographic run of the preceding
sample for optimal productivity and for highest system
utilization.
Automated liquid handling
in the MPS can be used for
additional sample prep steps. As part of the sample prep
procedure, standard addition can be performed automatically for
best possible analytical results.
Sample prep steps are
selected by mouse-click from a
pull-down menu using the PrepBuilder function of the GERSTEL
MAESTRO software.
A graphical scheduler
display for the DHS process
provides an overview of the time required for each individual
step and for the complete sequence of samples for easier
planning.

Just one method and one
sequence table controls the
complete process from DHS through sample prep and sample
introduction to GC/MS analysis. Integrated control ensures a
more efficient operation with less risk of error. The GERSTEL
MAESTRO Software operates stand-alone or fully integrated with
the Agilent Technologies ChemStation Software.
More info on DHS
SPEC:
DHS |